MODEL
CNX
VACUUM OVEN
-
Operating
Continuous Temperature Range: 100-482° F (30-250° C)
-
Operating
Vacuum Range: 760 to 1 Torr or 0.0 to 30.0-in Hg
-
Control
Accuracy +/- 2°C
or 1% of span whichever is greater
-
Rectangular
Chamber Type Construction
-
Standard
Platen Dimensions:
-
Model
36CNX-XR – 35.00-in W x 35.00-in D
-
Model
30CNX-XR – 29.50-in W x 29.50-in D
-
Model
24CNX-XR – 23.50-in W x 23.50-in D
-
Model
18CNX-XR – 17.50-in W x 17.50-in D
-
Model
12CNX-XR – 11.50-in W x 11.50-in D
-
Temperature
Uniformity (9-pt survey results) measured at 1-Torr
-
Model
36CNX-XR +/- 2.5°C
-
Model
30CNX-XR +/- 2.0°C
-
Model
24CNX-XR +/- 2.0°
C
-
Model
18CNX-XR +/- 2.0°
C
-
Model
12CNX-XR +/- 1.5°
C
The work chamber
is constructed of 304L or 316L stainless steel and features our unique
proprietary high strength gusseting system. The interior of the chamber is
fully welded which adds strength and improves its sealing capabilities. Chambers
are then passivated once all welding has been completed leaving a clean impurity
free finish. Other features include:
-
Proprietary
electrically transitioned aluminum or stainless steel platens machined true
and flat.
-
Our
Multilayer insulation technique is used on the rear wall where chamber
penetrations and ports are provided.
-
The vessel
door features our unique ultra heavy gusseting system, which offers anti
deflection capabilities during demanding thermal cycles. Vessel doors are
fully insulated using our multilayer type procedure maximizing efficiency,
reducing losses and thermal loading to the surrounding environment.
-
One-piece
constructed large diameter silicone door O-ring offers superior heat
resistance and sealing.
-
Vacuum levels
are achieved using simple single set-point direct acting solenoids, all
fully programmable from our F4AV controller. Accuracy is typically +/- 2
Torr during vacuum and +/- 1 Torr using a flow and pressure regulated
backfill source. A digital absolute vacuum explosion proof transducer
provides accurate and stable pressure level feedback to the F4AV
controller. Typical operating control range is expressed in Torr, 760 to 1,
but can be easily set-up for 0.0 to 30.0-inches of Mercury.
-
X-Purge
System designed for use in Class I / Division 1 hazardous locations
-
Sealed and
pressurized electrical housings on all electrical components including
heaters, control system and all motors.